User Tools

Site Tools


strutture:lnf:da:plasmonx:lpa:spectrometer:v1commissioning:jul2012

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
strutture:lnf:da:plasmonx:lpa:spectrometer:v1commissioning:jul2012 [2012/07/12 17:24] cgatti@infn.itstrutture:lnf:da:plasmonx:lpa:spectrometer:v1commissioning:jul2012 [2012/07/25 12:30] (current) faccini@infn.it
Line 1: Line 1:
 +=== July 10th 2012 ===
 +
 +
 +
 +
 +| Run | time | run conditions | comments |
 +| 760 | 12:30 | pedestals| ch 0,1,2 with PMTs|
 +| 761 | 12:40 | LED | HV on |
 +| 762 | 13:00 | LED | HV on, change timing|
 +| 763 | 18:35 | electrons  but not  to chamber | HV 900V|
 +| 764 | 19:15 | electrons  but not  to chamber | HV 500V|
 +| 765 | 19:45 | electrons  but not  to chamber | HV 500V, ch0 light tight, flanges darkened|
 +| 766 | 19:57 | electrons  but not  to chamber | HV 500V, ch0 light tight|
 +| 767 | 19:57 | electrons  but not  to chamber | HV 500V, ch0 light tight, dt=150|
 +| 768 | 19:57 | electrons  but not  to chamber | HV 500V, ch0 light tight, dt=250|
 +| 769 | 20:21 | electrons  but not  to chamber | HV 500V, ch0 light tight, dt=50, magnet on|
 +| 771-781 | 22:31 |problems after powercut| trigger timing changed |
 +| 782 | 0:21 |electrons in chamber|HV 500V, all light tight, dt=50, B=0 |
 +| 783 | 0:21 |electrons in chamber|HV 500V, all light tight, dt=50, B=0.45|
 +
 +=== July 11th 2012 ===
 +
 +HW changes:
 +  1) the second stage of the dual timer has been replaced by a Stanford for reliable and reproducible setting.
 +  2) all five channels have been equipped with PMT+Electronics
 +  3) the chamber of the spectrometer has been connected with the interaction region (i.e. the glass window removed)
 +  4) a lead collimator has been inserted between the gas jet and the spectrometer
 +  
 +Plan of the day:
 +  A) scan in integration time to find the best timing [with small magnet in]
 +  B) scan in HV to equalize gains [with small magnet in]
 +  C) spectrum with B=0 and B>0 (with small magnet out, having subtracted the common background)
 +  
 +  
 +
 +|Run | time | run conditions | comments |
 +|789 | 15:47 |No Beam HV on ch 0-4 500 V | pedestals|
 +|790 | 16:22 |No Beam HV on ch 0-4 900 V | pedestals|
 +|793 | 16:22 |electrons  | HV=500V. EM Noise is Back!|
 +|794-797| 17-18| Trigger Test|the trigger doesn't works|
 +|798| 19:38|T (signal splitting) removed from trigger |the trigger works|
 +|799|19:44|HV = 0V, laser+plasma|Electronic noise|
 +|800|20:00| HV=0, laser+plasma, reinforce grounding|
 +|801|20:04| try HV=500 but impossible to take data|
 +|804|22:50| HV=700 PMT 0, HV=0 PMT 1-4|saturation on PMT 0|
 +|805|22:52| HV=600 PMT 0 1 2, HV=0 PMT 3 4|now verify that noise on PMT0 increases with num of connected cards|
 +|806|23:26| HV=600 PMT 0 1 2, HV=0 PMT 3 4|only card 1 is connected, with the attenuator|
 +|807|23:36| HV=600 PMT 0 1 2, HV=0 PMT 3 4| B=0.45T|
 +|808|23:47| HV=600 PMT 0 1 2, HV=0 PMT 3 4| B=0.45T, card 1 2 3 are connected with attenuators|
 +|809-814| 23:59|HV=600 PMT 0 1 2, HV=0 PMT 3 4| B=0, scan in time reveals that t=0 is the best -->anticipate trigger|
 +|816-821| 24:10|HV=600 PMT 0 1 2, HV=0 PMT 3 4| B=0, scan in time: best timing dt = 100 ns|
 +|822|24:27|HV=550 V PMT 0, HV=600 PMT 1 2, HV=0 PMT 3 4| B=0|
 +|823|24:30|HV=550 V PMT 0, HV=600 PMT 1 2, HV=0 PMT 3 4| B=0|
 +|824|24:37|HV=500 V PMT 0, HV=600 PMT 1 2, HV=0 PMT 3 4| B=0|
 +|825|24:37|HV=OFF| B=0|
 +|826|24:41|HV=900 PMT 0 1 2| B=0|
 +|827|24:45|HV=900 PMT 0 1 2, dt=0| B=0|
 +
 +=== July 12th 2012 ===
 +Investigated noise problem. Disappears by using screened cable for DY12
 +
 +|Run | time | run conditions | comments |
 +| 829|13:29 |hv 800 ch 0,1,2 |B=0 |
 +| 830|13:32 |hv 900 ch 0,1,2 |B=0 |
 +| 831|13:35 |hv 800 ch 0,1,2 | B=0 magnete flame in|
 +| 832|14:10   |hv 500 ch 0,1,2 | B=0 magnete flame in remove attenuators|
 +| 833|14:15   |hv 600 ch 0,1,2 | B=0 magnete flame in remove attenuators|
 +| 834-852|15:33   |hv 600 ch 0-4 | trigger problems|
 +| 853-858|16:40   |hv 600 ch 0-4 | time scan|
 +| 859-870|16:45   |hv 500 ch 0-4 | time scan|
 +| 871-872|17:30   |HV scan to equalize| establish that HV={580,550,500,480,480}. Found that last two cables are swapped |
 +| 873 | B>0| turn on magnet. Hv default.| Elettroni deviati!!!|
 +| 874 | B>0 | magnet ON. HV default |  costanti di calibrazione Offline C=(1.2,1,1.5,1.1,1.4)|
 +| 875 | B>0 | HV def | Collimatore: 2 blocchetti 5 cm Pb dopo la finestra con 2 mm separazione: 40 mrad * 100 cm --> dx=4 cm |
 +| 876 | B=0 | HV def  | Collimatore; B ramping down? | 
 +| 877 | B=0 | HV def  | Collimatore | 
 +| 878 | B=0 | HV def  | Collimatore | 
 +| 879 | B=0 | HV def  | Collimatore chiuso; scambiati canali 3 e 4 che erano invertiti; C=(1.2,1,1.5,1.4,1.1) | 
 +| 882 | B=0 | HV def  | Collimatore chiuso | 
 +| 883 | B=0 | HV def  | Collimatore chiuso; Flame Magnet IN; Use for Ped File | 
 +| 884 | B=0 | HV def  | Collimatore aperto 2mm; Flame Magnet OUT; usb problem at the end| 
 +| 885 | B=0 | HV def  | Collimatore aperto 2mm; Flame Magnet OUT| 
 +| 886 | B=0 | HV def  | Collimatore aperto 1cm; Flame Magnet OUT| 
 +| 887 | B=0.5 T | HV def  | Collimatore aperto 1cm; Flame Magnet OUT| 
 +=== July 13th 2012 ===
 +
 +Debug of spectrometer trigger. Problems with upper part of logic unit, replaced with lower part. Dual timer removed, veto implemented with Stanford D signal set to 500 mus. Delay and width of trigger button revised, set to 500 mus. Missing alignment with photocatode.
 +
 +
 +=== July 18th 2012 ===
 +Pedestal studies
 +
 +| 896 | 12:21| HV off| Only maroc 0 board+cable; no laser|
 +| 897 | 12:26| HV off| Only maroc 0 board+cable; no laser| RMS 12 cts |
 +| 899 | 12:26| HV off| Only maroc 0 board no cable; no laser; improved trigger cable| RMS 1.5 cts |
 +
 +Giulietto Felici works on electronics to improve grounding and box.
 +
 +=== July 25th 2012 ===
 +
 +test electronics with new shielding
 +
 +| 900 | 10:37| pedestals| 5cards but no cables; RMS 1-2 ADC|
 +| 901 | 10:40| pedestals| 5cards cable #0 in; HV off; RMS ~6 for first 43 channels, 10-18 for the others |
 +| 902 | 10:44| pedestals| 5cards cable #0 in; HV=900V; RMS ~6 for first 43 channels, 10-18 for the others |
 +| 903 | 10:52| LED run| 5cards cable #0 in; HV=900V |
 +| 904 | 14:26| pedestals| 5cards all cables in; HV=0V |
 +| 905 | 14:29| pedestals| 5cards all cables in; HV=default |
 +| 905 | 14:29| pedestals| 5cards all cables in; HV=default -- high stat |
  

Donate Powered by PHP Valid HTML5 Valid CSS Driven by DokuWiki