User Tools

Site Tools


strutture:lnf:da:btf:dispositivi:btf_remotabletable

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
strutture:lnf:da:btf:dispositivi:btf_remotabletable [2014/02/11 17:17] foggetta@infn.itstrutture:lnf:da:btf:dispositivi:btf_remotabletable [2025/01/10 09:40] (current) diociaiu@infn.it
Line 1: Line 1:
 +==== REMOTABLE BENCH ====
 +\\
 +{{:strutture:lnf:da:btf:dispositivi:impianti:tavolo.pdf|Manuale Tavolo Remoto - IT}}\\
 +{{:strutture:lnf:da:btf:dispositivi:remotetrolleytable_manual_en.pdf| Remote Trolley Bench Manual - EN}}
 +\\
 +
 +
 +**-> Coordinates**
 +The beam Y position w.r.t. the floor is 1240 mm
 +\\
 +{{:strutture:lnf:da:btf:dispositivi:impianti:tavolo1.jpg|Bench from the top}}
 +{{:strutture:lnf:da:btf:dispositivi:impianti:tavolocontr.jpg|VI table}}\\
 +
 +    Movable platform X AXIS => 
 +      + means toward bench's cabinet
 +      - means the viceversa
 +    Movable platform Y AXIS =>
 +      + means up
 +      - means down
 +
 +  *The unit is millimeters
 +  *Platform Homing position
 +    * HORIZONTAL = fully opposite the bench cabinet
 +    * VERTICAL = fully down
 +  *The absolute position is relative to software zero or homing position
 +
 +**-> User Menu**
 +
 +   1. Before installing please read the manual (on the top of this page) and put the table in homing position!
 +   2. When you setup the software limits, in case of successive advance menu opening, you will reset the user zero position
 +   3. The software limits are to be intended as the modulus of the maximum differential movement from the user zero position
 +
 +**-> Zero position and set up the software limits**
 +
 +   1. Get your own bench platform position
 +   2. setup the xy limits and press setup button (now the X and Y position is set to 0)
 +   3. free to move the table within new limits
 +   
 +   ATTENTION!!! If you want to reset the XY limits, you HAVE TO redo all the point 1., 2., 3.
 +
 +**-> Movable platform Minimum step = 0.5 mm**\\   
 +**-> Movable platform Maximum step = 50 mm**\\
 +**-> Movable platform Recursive precision = 0.2 mm (relative to the payload)**\\
 +
 +**-> Maximum range **\\
 +   1. X = 984.0 mm
 +      Y = 380.0 mm (software 335.0 mm)
 +      
 +      
 +
 +**-> Beam-based alignment within the ground reference position**
 + \\
 +    X0=492mm (middle of frame) Y0 = 330.0 mm
 + \\
 +\\
 +\\ **Features and dimensions of the bench**
 + 
 +  * Overall dimensions: 2100 x 705 x 915 mm (with the platform in the lowest position).
 +  * Weight of the bench: about 250 daN (including all the electrical components attached).
 +  * Travel of the moving platform:
 +  *   * X-axis = 984 mm,
 +  *   * Y-axis = lowest position 915 mm (ground clearance). Highest position (again from the ground) 1250 mm; 335 mm maximum excursion.
 +  * Maximum lifting weight load on the platform: 200 daN (kg).
 +  * Bench wheels for moving: n.4. Two fixed and two pivoting for manual handling, load capacity 180 daN (each).
 +  * Handles: n.4. Two per side for taking and moving the bench, load capacity each. 200 daN. (IMPORTANT: not to be used for lifting).
 +  * Feet: n.4. Use feet for supporting, position fixing, offset adjustment and leveling to the floor. Maximum load 2,300 daN.
 +  * Steel brackets provided with an eyelet for attaching lifting cables: n.4.
 +  * Movement Protections (automatic roller blinds for X-axis movement): 2.
 +  * Accident prevention spacers (Y-axis movement): 4.
 +\\
 +
 +**Iron Platform Holding Surface**\\
 +\\
 +  * The movable surface has 25 drilled holes compatible with M8 screw. \\
 +  * The holes are placed in a grid having steps of 10cm, starting 10 cm from borders.\\
 +  * The thickness of the surface is 2 cm.\\
 +  * Iron platform: 600 x 600 x 20 mm.
 +
 +\\
 +\\
 +{{:strutture:lnf:da:btf:dispositivi:table_Jan24.png?200|}}
 +{{:strutture:lnf:da:btf:immagine_whatsapp_2025-01-10_ore_10.36.34_01444ad4.jpg|}}
 +{{:strutture:lnf:da:btf:immagine_whatsapp_2025-01-10_ore_10.36.34_01444ad4.jpg|}}\\
 +\\
 +\\
 +
 +**Troubleshoot**\\
 +In case of power shutdown, the table will lose the homing point.
 +
 +Se si cambia l'escursione massima senza premere “Set” il sistema non ricentra lo zero utente (mantiene lo zero utente precedente) ma rileva in maniera errata i nuovi limiti di escursione massima
  

Donate Powered by PHP Valid HTML5 Valid CSS Driven by DokuWiki