+ | ==== WCM ==== | ||
+ | |||
+ | //OUTSIDE BTF LINE//\\ | ||
+ | Two WCM are placed before the pulsed magnet (WCMTM01) and after SLTB001 (WCMTB01)\\ | ||
+ | One WCM is installed at the end of the DHSTB002 dipole straight line (WCMTB02).\\ | ||
+ | |||
+ | **BERGOZ ICT 122-070-05-1**\\ | ||
+ | [[http:// | ||
+ | {{: | ||
+ | \\ | ||
+ | |||
+ | **Current to charge conversion with QDC:** | ||
+ | |||
+ | After having performed a pedestal subtraction, | ||
+ | |||
+ | ADC(pede sub)/ | ||
+ | |||
+ | where:\\ | ||
+ | * ADC(pede sub) = pedestal subtracted ADC value | ||
+ | * FS(ch)= range in [pC] of the selected dual range channel | ||
+ | * Q is the charge expressed in pC | ||
+ | * G=10(WCM btf), 20(WCM linac) | ||
+ | |||
+ | **Current to charge conversion with One electron:** | ||
+ | |||
+ | Put on peak the value of | ||
+ | * 0.819 for a 800pC | ||
+ | * 0.728 for 900pC | ||
+ | * 6.552 for 100pC | ||
+ | |||
+ | |||
+ | and validate,\\ | ||
+ | \\ | ||
+ | the #ele control will measure the value of total integrated charge in unit of 10^7 electrons\\ | ||
+ | |||
+ | **Current to charge conversion with oscilloscope: | ||
+ | |||
+ | -- Integrated Charge with measure reference bars (measure outputted by scope)= C [pV*s] | ||
+ | -- Pedestal = F[pV*s] (same as signal measure, triggered with DHSTB001 in stby mode) | ||
+ | -- Internal resistance = R[Ohm] | ||
+ | -- Bergoz monitor calibration => Q(beam)/ | ||
+ | -- Bunch pulse width = dt [ns] | ||
+ | |||
+ | Formula to obtain induced charge on WCM via scope: | ||
+ | Q[pC]=g*[(C-F)/ | ||
+ | |||
+ | Formula to obtain mean current from single bunches (To be compared with LINAC WCMTM001 in order to measure transport efficiency): | ||
+ | I[mA]= | ||
+ | | ||
+ | | ||
+ | Measure 20131213 on WCMTB002 (straight line) with pedestal subtracted (201312101258.ped) of value (63.02/ | ||
+ | |||