 +====== Faraday Cup ======
 +
 +Adone Faraday cup\\
 +dimensions\\
 +copper shielding\\
 +
 +
 +**Current to charge conversion with QDC Caen V965:**
 +
 +After having performed a pedestal subtracion, the ADC value is converted to charge with this formula:\\
 +
 +ADC(pede sub)/4095*FS(ch)= Q [pC]\\
 +
 +where:\\
 +  * ADC(pede sub) = pedestal subtracted ADC value
 +  * FS(ch)= range in [pC] of the selected dual range channel  (100pC or 900pC, new version of V965)
 +  * Q is the charge expressed in pC
 +
 +**Current to charge conversion with One electron:**
 +
 +Put on peak the value of 
 +  * 6.006 for 900pC
 +  * 54.054 for 100pC
 +
 +
 +and validate,\\
 +\\
 +the #ele control will measure the value of total integrated charge in unit of 10^7 electrons\\
 +
 +**Current to charge conversion with oscilloscope:**
 +
 +  -- Integrated Charge with measure reference bars (measure outputted by scope)= C [pV*s]
 +  -- Pedestal = F[pV*s] (same as signal measure, triggered with DHSTB001 in stby mode)
 +  -- Internal resistance = R[Ohm]
 +  -- Bergoz monitor calibration => Q(beam)/Q(out) @ 50 Ohm 
 +  -- Bunch pulse width = dt [ns]
 +
 +Formula to obtain iduced charge on WCM via scope:
 +  Q[pC]=[(C-F)/R]
 +
 +Formula to obtain mean current from single bunches (To be compared with LINAC WCMTM001 in order to measure transport efficiency):
 +  I[mA]=  { (C[pVs]-F[pVs]) /50[Ohm]}/dt[ns]
 +    
 +  
 + 
  