strutture:lnf:da:btf:dispositivi:btf_diagnostic:faraday_cup
Differences
This shows you the differences between two versions of the page.
strutture:lnf:da:btf:dispositivi:btf_diagnostic:faraday_cup [2022/07/11 15:12] – created foggetta@infn.it | strutture:lnf:da:btf:dispositivi:btf_diagnostic:faraday_cup [2022/07/11 15:14] (current) – foggetta@infn.it | ||
---|---|---|---|
Line 1: | Line 1: | ||
+ | ====== Faraday Cup ====== | ||
+ | |||
+ | Adone Faraday cup\\ | ||
+ | dimensions\\ | ||
+ | copper shielding\\ | ||
+ | |||
+ | |||
+ | **Current to charge conversion with QDC Caen V965:** | ||
+ | |||
+ | After having performed a pedestal subtracion, the ADC value is converted to charge with this formula:\\ | ||
+ | |||
+ | ADC(pede sub)/ | ||
+ | |||
+ | where:\\ | ||
+ | * ADC(pede sub) = pedestal subtracted ADC value | ||
+ | * FS(ch)= range in [pC] of the selected dual range channel | ||
+ | * Q is the charge expressed in pC | ||
+ | |||
+ | **Current to charge conversion with One electron:** | ||
+ | |||
+ | Put on peak the value of | ||
+ | * 6.006 for 900pC | ||
+ | * 54.054 for 100pC | ||
+ | |||
+ | |||
+ | and validate,\\ | ||
+ | \\ | ||
+ | the #ele control will measure the value of total integrated charge in unit of 10^7 electrons\\ | ||
+ | |||
+ | **Current to charge conversion with oscilloscope: | ||
+ | |||
+ | -- Integrated Charge with measure reference bars (measure outputted by scope)= C [pV*s] | ||
+ | -- Pedestal = F[pV*s] (same as signal measure, triggered with DHSTB001 in stby mode) | ||
+ | -- Internal resistance = R[Ohm] | ||
+ | -- Bergoz monitor calibration => Q(beam)/ | ||
+ | -- Bunch pulse width = dt [ns] | ||
+ | |||
+ | Formula to obtain iduced charge on WCM via scope: | ||
+ | Q[pC]=[(C-F)/ | ||
+ | |||
+ | Formula to obtain mean current from single bunches (To be compared with LINAC WCMTM001 in order to measure transport efficiency): | ||
+ | I[mA]= | ||
+ | | ||
+ | | ||
+ | |||