CALOBTF1


Lead Glass → Cherenkov

From the former OPAL electromagnetic calorimeter

Referred measuring System
Caen V965 - CH4 900pC (old Type)
Caen N625 - CH2 @ -7mV bias


N=184.85*(x)^7.705*Ebeam\E0
N = single particle value [adc unit, pedestal subtracted]
x = V/1200 [V]
V= Voltage on CALO (Minimum Voltage 450V)
Ebeam = Energy beam [MeV]
E0 = 447 [MeV]

TABLE @ 447 MeV
Last calibration @450MeV 20220505

V N lastcal N m=1 sigma lastcal
1300 360.9 350.6 41.8
1200 186.75 188 22.2
1100 92.51 96.1 11,4
1000 42.54 45.3 5.6
900 18.03 19.6 2.58
800 6.903 8.23
700 2.325 2.78
600 0.662 0.84
500 0.1507 0.216
450 0.092
400 0.0415

Oscilloscope to Multiplicity calibration
To recover multiplicity by your own measurement system with our CALOBTF1 signal, follow these steps:

S=(GS-P)/(50)= [V*s/Ohm]
M=S/900E-12*4096/N = [C*ADC/C/ADC]

CALO as BACKGROUND gamma detector for HIGH INTENSITY operation on N@BTF
We can use CALOBTF1 as background gamma detector to discover the upper limit of gamma to prevent the safety sistem to trips.
We put CALOBTF1 in the bent middle line at 3.8 meters from exit bent pipe with 500V supply
Normal operation means up to 20mV peak to peak signal, if above the safety trips