Lead Glass → Cherenkov
From the former OPAL electromagnetic calorimeter
Referred measuring System
Caen V965 - CH4 900pC (new Type)
Caen N625 - CH2 @ -7mV bias
N=[239.7*(x)^7.705]*Ebeam\E0
N = single particle value [adc unit, pedestal subtracted]
x = V/1200 [V]
V= Voltage on CALO (Minimum Voltage 450V)
Ebeam = Energy beam [MeV]
E0 = 450 [MeV]
TABLE @ 450 MeV
Last calibration @450MeV 20220505
V | N |
---|---|
1300 | 444.14 |
1200 | 239.7 |
1100 | 122.6 |
1000 | 58.83 |
900 | 26.122 |
800 | 10.540 |
700 | 3.768 |
600 | 1.1496 |
500 | 0.28288 |
Oscilloscope to Multiplicity calibration
To recover multiplicity by your own measurement system with our CALOBTF signal, follow these steps:
S=(GS-P)/(50)= [V*s/Ohm]
M=(S/900E-12)*4096/N = [(C/C)*(ADC/ADC)]
CALO as BACKGROUND gamma detector for HIGH INTENSITY operation on N@BTF
We can use CALOBTF as background gamma detector to discover the upper limit of gamma to prevent the safety sistem to trips.
We put CALOBTF in the bent middle line at 3.8 meters from exit bent pipe with 500V supply
Normal operation means up to 20mV peak to peak signal, if above the safety trips